Acta Scientific Medical Sciences (ASMS)(ISSN: 2582-0931)

Research Article Volume 9 Issue 10

Various Types of Coating of the Mechanical Circulation Support Pump Impact on Hemolytic and Homeostatic Blood Parameters

KO Golovin1,2, AM Golovin2 and RI Aizman1,3*

1Novosibirsk State Pedagogical University, Novosibirsk, 630126, Vilyuiskaya Str., 28, Russia
2NPK "IMPULSE-Project", Novosibirsk, Russian Federation 630073, Novosibirsk, Gorsky Microdistrict, 10, Russia
3Novosibirsk Scientific Research Institute of Hygiene of Rospotrebnadzor, 630108, Novosibirsk, Parkhomenko Str., 7, Russia

*Corresponding Author: RI Aizman, Novosibirsk State Pedagogical University, Novosibirsk, 630126, Vilyuiskaya Str., 28, Russia.

Received: February 17, 2025; Published: September 11, 2025

Abstract

Mechanical circulatory support can lead to hemolysis, thrombosis, bleeding, and other complications. The aim of the study was to determine the degree of influence of various types of coating of the disk pump for mechanical circulatory support on hemolytic blood parameters.

Materials and Methods: In vitro studies were conducted on a specialized device that simulated a large circle of blood circulation. Human donated blood was used for the study.

Results and Discussion: It was found that using an uncoated and diamond-coated disc pump, an increase in the level of FreeHb in blood plasma was noted from a baseline value of 0.01 g/l to 0.12 g/l and from 0.02 to 0.09 g/l, respectively. In studies, where a disk pump with a surface treated by electron beam with a silicon-carbon coated was used, the concentration of FreeHb in blood plasma remained at the background level.

Conclusion: It is shown that a disk pump with a surface treated by electron beam with a silicon-carbon coated caused the smallest changes of the studied blood parameters.

 Keywords: Disc Pump; Coating; FreeHb; Hemolysis

References

  1. Slaughter MS., et al. “Heart-Mate II Clinical Investigators. Clinical management of continuous-flow left ventricular assist devices in advanced heart failure”. The Journal of Heart and Lung Transplantation 4 (2010): S1-39.
  2. Birschmann I., et al. “Ambient hemolysis and activation of coagulation is different between HeartMate II and Heart Ware left ventricular assist devices”. The Journal of Heart and Lung Transplantation1 (2014): 80-87.
  3. Kirklin JK., et al. “Seventh INTERMACS annual report: 15,000 patients and counting”. The Journal of Heart and Lung Transplantation12 (2015): 1495-1504.
  4. Vieira JL., et al. “Mechanical circulatory support devices in ad-vanced heart failure: 2020 and beyond”. Progress in Cardiovascular Diseases5 (2020): 630-639.
  5. Papanastasiou CA., et al. “Comprehensive review of hemolysis in ventricular assist devices”. World Journal of Cardiology 7 (2020): 334-341.
  6. Itkin GP., et al. “Special features of long-term mechanical circulatory support using continuous flow pumps”. Bulletin of Transplantology and Artificial Organisation14 (2014): 110-115.
  7. Zinoviev R., et al. “In full flow: left ventricular assist device infections in the modern era”. Open Forum Infectious Diseases 7 (2020): ofaa124.
  8. Takaseya T., et al. “In vivo biocompatibility evaluation of a new resilient, hard-carbon, thin-film coating for ventricular assist devices”. Artificial Organs12 (2010): 1158-1163.
  9. Mueller MR., et al. “In vitro hematological testing of rotary blood pumps; remarks on standardization and data interpretation”. Artificial Organs2 (1993): 103-110.
  10. Nakazawa T., et al. “Development and initial testing of a permanently implantable centrifugal pump”. Artificial Organs7 (1997): 597-601.
  11. Xuan Y., et al. “Exploration of hemolytic model of axial blood pump”. Journal of Experimental and Integrative Medicine4 (2012): 365-371.
  12. Wright G. “Haemolysis during cardiopulmonary bypass: update”. Perfusion 5 (2001): 345-351.
  13. Ravichandran AK., et al. “Hemolysis in left ventricular assist device: a retrospective analysis of out-comes”. The Journal of Heart and Lung Transplantation1 (2014): 44-50.
  14. Dmitrieva OYu., et al. “Hemolysis studies of an implantable axial pump for two-stage heart transplantation in children”. Bulletin of Transplantology and Artificial Organisation19 (2017): 22-27.
  15. Bourque K., et al. “Design rationale and preclinical evaluation of the HeartMate 3 Left Ventricular Assist System for hemocompatibility”. ASAIO Journal4 (2016): 375-383.

Citation

Citation: RI Aizman., et al. “Various Types of Coating of the Mechanical Circulation Support Pump Impact on Hemolytic and Homeostatic Blood Parameters ”.Acta Scientific Medical Sciences 9.10 (2025): 39-44.

Copyright

Copyright: © 2025 RI Aizman., et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.




Metrics

Acceptance rate30%
Acceptance to publication20-30 days
Impact Factor1.403

Indexed In





Contact US