Acta Scientific Paediatrics (ISSN: 2581-883X)

Research ArticleVolume 4 Issue 3

Development of a Portable Non-Invasive Oral Motor Assessment Device and its Reliability

Payam Seyedkalateh1, Elnaz Pooralijani2, Seyed Ahmad Hosseini3*, Peyman Asghari4 and Mehdi Nouri5

1Speech and Language Pathologist, Sayyad Shirazi Hospital, Gorgan, Iran
2Master Student of Counseling in Midwifery, Gorgan University of Medical Science, Gorgan, Iran
3Child Neurologis, Assistant Professor of Pediatric Department, Neonatal and Children's Health Research Center Taleghani Medicine Educational Center Golestan University of Medical Sciences Gorgan Iran
4Electronics Engineer, Gorgan, Iran
5Phd Student of Educational Psychology, Ferdowsi University of Mashhad, Mashhad, Iran

*Corresponding Author: ed Ahmad Hosseini, Child Neurologis, Assistant Professor of Pediatric Department, Neonatal and Children's Health Research Center Taleghani Medicine Educational Center Golestan University of Medical Sciences Gorgan Iran

Received: January 25, 2021; Published: February 15, 2021

Citation: Payam Seyedkalateh., et al. “Development of a Portable Non-Invasive Oral Motor Assessment Device and its Reliability”. Acta Scientific Paediatrics 4.3 (2021): 50-55.

Graphical Abstract

Figure

Abstract

Purpose : Development of a portable non-invasive oral motor assessment device.

Method: New device is designed and twenty people (11 male and 9 female) in deferent range of age with no history of swallowing, orofacial, gastrointestinal, respiratory or neurologic problem were selected. sample was evaluated twice in 15 minutes intervals with device for assessment of reliability.

Result: There are correlation between the two measurements in test-retest and its significance and the degree of reliability of each measurement.

Conclusion: It can be concluded that these tools have an acceptable reliability and reliability index and it can be used as a reliable instrument in the evaluation of some oral motor skills.

Keywords: Oral Motor; Infant Feeding; Sucking Assessment

Introduction

  The craniofacial and masticatory musculature is deployed for a variety of behaviors, including speech, communicative and noncommunicative facial gestures, biting, chewing, swallowing, licking and ventilation. These diverse behaviors use many of the same muscles but with differing patterns of activation [1]. Oral-motor skills refer to movement of the muscles of face (e.g. mouth, jaw, tongue and lips). This includes muscle tone, muscle strength, range of motion, speed, oordination and dissociation (the ability to move oral structures, such as the tongue and lip, independently of each other) [2,3]. Clinical experience suggests that the acquisition and maturation of oral-motor movements underlies sound production and feeding skills (e.g. sucking, biting and chewing) [4]. Children with developmental disabilities may demonstrate oral-motor patterns that are not observed in typical development, atypical oral-motor patterns include jaw thrusting, tongue thrust, tonic bite reflex, lip retraction, tongue retraction and nasal regurgitation [5].

A child oral-motor skills may be assessed speech language pathologist.

  Given the large number of clients in speech therapy consist of patients with impaired oral motor skills (adults and children), there are several device [6-12], assessment tests designed to assess the motor skills for children’s [13-15] and adults [8]. But according to the researchers' knowledge of this study, there is no new device for assessing the oral motor skills. In this research we are developed the new portable device that can assessment the some of oral motor skills (such as blowing, nutritive sucking, non-nutritive sucking, finger feeding sucking, nipple feeding sucking, tongue palate pressure, up down of larynx in swallowing and lip power) and We examined its reliability in this study. The device registered with No.101149 in the Iranian Patent Organization.

Materials and Methods

Subjects

The following formula was used to calculate the sample size of the research:

  According to the variance of the variables of this study (S2 = 38) and the volume of the effect (d = 2.7) at a significance level of 95%, the sample size of 20 people was obtained.

Twenty people (11 male and 9 female) in deferent range of age with no history of swallowing, orofacial, gastrointestinal, respiratory or neurologic problem were selected.

Device

  The schematic of the device is shown in figure 1. As shown, it is a portable device that has two sensors, an internal computer, an internal processor, printers, and special probes for evaluation.

Figure 1

  The sensor 1 is MPXV7007 series piezoresistive transducers are state-of-the-art monolithic silicon pressure sensors. The sensor 2 is piezoelectric sensor (Shown in figure 1) is the kind of pressure sensors used [16]. It is made by solid materials, which can accumulate electric charges in response to applied mechanical stress.

  Reliability is the degree of uniformity of results over a given time and under the same conditions and with the same working method as is characterized by the reliability and repeatability of the measured results [17]. In this method first explained the process of working to people. Then, each sample was evaluated twice in 15 minutes intervals. The problem of determining the appropriate time interval for test run and the test takers' reluctance to run second run has been discussed by many scholars. According to the researchers, three types of effects, including fatigue effect, memory effect, Genuine Change Effect, will be effective in maintaining test retention value [17]. This test may only produce the effects of partial fatigue from the production of speech tests, which is a 15-minute gap between the test and the retest and second tests.

Test-retest of sensor 1 (Suction section)

  As seen in figure 2, the suction tube connects to the device from one side to the mouth tube at the other. The tester then inserts the tube into the case's mouth and after ensuring that the case is positioned correctly, he wants to suck (at least 3 seconds).

Figure 2

  As the case starts sucking, the tester puts the device in recording mode and, after sucking the case, automatically saves the numbers in the internal memory and also transfers them to the flash. After 15 minutes and sufficient rest, the above process is repeated again for retest.

Test-retest of sensor 1 (blowing section)

  As seen in figure 3, the blowing tube connect to the device from one end to the mouth tube at the other site. The tester then inserts the tube into the test case's mouth and after ensuring that the case is positioned correctly. The viewer wants to blow (at least 3 seconds). As the sucker begins to suck, the tester puts the device in recording mode and when the case is finished, the device automatically stores the numbers in the internal memory and also transfers them to the flash. After 15 minutes and sufficient rest, the above process is repeated again for retest.

Figure 3

Test-retest of sensor 2 (Lips closure pressure)

  As seen in figure 4, the tester insert the sensor between the lips and ensure that the sensor is positioned correctly. at the same time as the case's pressure, the tester puts the device in recording mode and upon completion of the case pressure, the device automatically stores the numbers in the internal memory and also transfers them to the flash. After 15 minutes and sufficient rest, the above process is repeated for again for retest.

Figure 4

Results

As seen in table 1, the mean negative pressures of suction in the test and retest are shown.

Retest

Test

Case

-0.910

-1.089

1

-10.211

-9.112

2

-12.759

-12.230

3

-13.89

-12.30

4

-7.019

-6.821

5

-11.630

-12.501

6

-5.807

-6.157

7

-14.924

-12.049

8

-10.914

-9.043

9

-4.30381

-4.53106

10

-1.126

-1.075

11

-3.091

-2.922

12

-17.725

-16.006

13

-2/578

-2/551

14

-2/995

-2/703

15

-13/117

-13/016

16

-3/06194

-2/9234

17

-1/795

-1/986

18

-22/060

-21/382

19

-13.283

-12/922

20

Table 1: Mean of sucking pressure.
Numbers are based on kilopascals(Kpa)

As seen in table 2, the mean blowing positive pressures in the test and retest are presented.

Retest

Test

Case

1.156

1.884

1

2.254

2.824

2

7.914

7.647

3

10.398

10.076

4

2.443

2.562

5

3.090

3.95

6

5.457

5.916

7

2.058

1.998

8

6.650

7.597

9

7.739

7.921

10

1.902

2.376

11

3.755

3.676

12

5.434

6.248

13

6/512

6/622

14

3/309

3/067

15

8/697

8/151

16

3/549

3/437

17

3/437

3/550

18

3/544

3/262

19

4/159

3/783

20

Table 2: Mean of blowing pressure.
Numbers are based on kilopascals(Kpa)

In table 3, the mean tongue and palate pressure are presented in the test and retest.

Retest

Test

Case

7.684

8.823

1

10.214

9.111

2

5.03695

9.45487

3

47.817

48.742

4

7.420

5.560

5

3.713

3.068

6

32.981

34.389

7

2.358

4.605

8

28.394

35.539

9

36.69

33.39

10

3.410

4.246

11

7.949

9.449

12

32.567

34.604

13

4.514

3.217

14

5.327

4.124

15

28.234

26.410

16

4.536

5.792

17

11.612

9.142

18

6.324

5.126

19

9.362

7.580

20

Table 3: Mean of lips closure pressure.
Numbers are based on kilopascals(Kpa)

  The reliability of the measuring instrument for information of sucking pressure, blowing pressure and lips closure pressure based on data obtained from 20 individuals according to table 4 is as follows.

Reliability

Sig. (2-tailed)

Pearson Correlation

Measures

0.995

0.000

0.991

Mean of sucking pressure

0.991

0.000

0.984

Mean of blowing pressure

0.992

0.000

0.985

Mean of lips closure pressure

Table 4
Numbers are based on kilopascals(Kpa)

  Table shows the correlation between the two measurements in Test-Retest, and its significance, and the degree of reliability of each measurement. The Spearman-Brown formula was used to calculate the reliability.

  According to the calculated reliability values, it can be concluded that these tools have an acceptable reliability and reliability index.

Discussion

  Digital measurement of air pressure is a continuous chain. That is, due to the changeable nature of air pressure, pressure sensors are forced to record all changes in air pressure within a defined time frame (according to the loaded schedule) in this type of system. The tester should select and analyze the range that is the most significant change in pressure related to tasks.

  According to the results obtained from sensor number one, it can be said that this sensor has good reliability and it can be used for studies such as infant nipple feeding assessment, infant finger feeding assessment, infant bottle feeding assessment, sucking assessment in adult and child’s, blowing assessment in adult and child’s.

  Sensor number two show good reliability. But the placement of the sensor and the place of pressure on the sensor are important. It is practically impossible to repeat the same test conditions for this sensor when do retesting. also due to the lightness and shape of the sensor, we can use it to receive online feedback in treatment.

Conclusion

  However, according to the obtained results, it can be suggested that if more comprehensive research is done using this device, it can be used as a reliable instrument in the evaluation of some oral motor skills.

Conflict of Interest

The authors report no declarations of interest.

Bibliography

  1. Kent RD. “Nonspeech oral movements and oral motor disorders: A narrative review”. American Journal of Speech-Language Pathology4 (2015): 763-789.
  2. Clark GT and Ram S. “Four oral motor disorders: bruxism, dystonia, dyskinesia and drug-induced dystonic extrapyramidal reactions”. Dental Clinics1 (2007): 225-243.
  3. Kumin L. “Resource guide to oral motor skill difficulties in children with down syndrome”. Loyal College of Meriland (2015).
  4. Arvedson JC., et al. “Pediatric swallowing and feeding: Assessment and management” (2019).
  5. Morris SE. “Pre-feeding skills. A comprehensive resource for mealtime development” (2000): 81-82.
  6. https://iopimedical.com/
  7. https://www.swallowsolutions.com/product-information/swallowstrong-device
  8. Nascimento MD., et al. “Reliability of the S-FLEX® device to measure non-nutritive sucking pressure in newborns”. Audiology-Communication Research (2019): 24.
  9. Chen L., et al. “A novel system to measure infants’ nutritive sucking during breastfeeding: the breastfeeding diagnostic device (BDD)”. IEEE Journal of Translational Engineering in Health and Medicine 6 (2018): 1-8.
  10. Chen CT., et al. “Wireless monitoring system for oral-feeding evaluation of preterm infants”. IEEE Transactions on Biomedical Circuits and Systems 5 (2015): 678-685.
  11. Taffoni F., et al. “Ecological sucking monitoring of newborns”. IEEE Sensors Journal 11 (2013): 4561-4568.
  12. Tamilia E., et al. “An automated system for the analysis of newborns’ oral-motor behavior”. IEEE Transactions on Neural Systems and Rehabilitation Engineering 12 (2015): 1294-1303.
  13. Skuse D., et al. “Schedule for oral-motor assessment (SOMA): methods of validation”. Dysphagia 3 (1995): 192-202.
  14. Palmer MM., et al. “Neonatal Oral-Motor Assessment scale: a reliability study”. Journal of Perinatology: Official Journal of the California Perinatal Association1 (1993): 28-35.
  15. De Oliveira Lira Ortega A., et al. “Assessment scale of the oral motor performance of children and adolescents with neurological damages”. Journal of Oral Rehabilitation9 (2009): 653-659.
  16. Selçuk B., et al. “Effect of temperature on electrophysiological parameters of swallowing”. Journal of Rehabilitation Research and Development3 (2007): 373.
  17. Mohammadbeigi A., et al. “Validity and reliability of the instruments and types of measurements in health applied researches”. Journal of Rafsanjan University of Medical Sciences12 (2015): 1153-1170.

Copyright: © 2021 Payam Seyedkalateh., et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.



News and Events


  • Certification for Review
    Acta Scientific certifies the Editors/reviewers for their review done towards the assigned articles of the respective journals.
  • Submission Timeline for Upcoming Issue
    The last date for submission of articles for regular Issues is April 30th, 2024.
  • Publication Certificate
    Authors will be issued a "Publication Certificate" as a mark of appreciation for publishing their work.
  • Best Article of the Issue
    The Editors will elect one Best Article after each issue release. The authors of this article will be provided with a certificate of "Best Article of the Issue"
  • Welcoming Article Submission
    Acta Scientific delightfully welcomes active researchers for submission of articles towards the upcoming issue of respective journals.

Contact US